Ambient air quality monitoring in India compares outdoor air with the National Ambient Air Quality Standards that CPCB notified on 18 November 2009 (NAAQS 2009). They set limits for 12 pollutants by averaging time and area category. CPCB's monitoring programme samples for 24 hours twice a week, giving 104 results a year.
This guide is for environmental labs, field teams and quality managers who run ambient air monitoring for industries, projects and regulators. It sets out the limits, the CPCB ambient air quality monitoring guidelines on frequency and siting, a worked PM10 calculation, and what a report should contain.
What are the NAAQS 2009 standards?
NAAQS 2009 is CPCB Notification No. B-29016/20/90/PCI-L dated 18 November 2009, issued under Section 16(2)(h) of the Air (Prevention and Control of Pollution) Act, 1981. It replaced the earlier standards of 1994 and 1998 and took effect immediately.
The notification has two area categories: "Industrial, Residential, Rural and Other Area", and "Ecologically Sensitive Area (notified by Central Government)". For most pollutants the two limits are the same; they differ only for the annual SO2 and NO2 standards.
| Pollutant | Averaging time | Industrial, residential, rural and other | Ecologically sensitive | Methods named in the notification |
|---|---|---|---|---|
| SO2, µg/m³ | Annual | 50 | 20 | Improved West and Gaeke; ultraviolet fluorescence |
| SO2, µg/m³ | 24 hours | 80 | 80 | As above |
| NO2, µg/m³ | Annual | 40 | 30 | Modified Jacob and Hochheiser (Na-arsenite); chemiluminescence |
| NO2, µg/m³ | 24 hours | 80 | 80 | As above |
| PM10, µg/m³ | Annual | 60 | 60 | Gravimetric; TEOM; beta attenuation |
| PM10, µg/m³ | 24 hours | 100 | 100 | As above |
| PM2.5, µg/m³ | Annual | 40 | 40 | Gravimetric; TEOM; beta attenuation |
| PM2.5, µg/m³ | 24 hours | 60 | 60 | As above |
| Ozone, µg/m³ | 8 hours | 100 | 100 | UV photometric; chemiluminescence; chemical method |
| Ozone, µg/m³ | 1 hour | 180 | 180 | As above |
| Lead, µg/m³ | Annual | 0.50 | 0.50 | AAS or ICP after sampling on EPM 2000 or equivalent filter; ED-XRF using Teflon filter |
| Lead, µg/m³ | 24 hours | 1.0 | 1.0 | As above |
| CO, mg/m³ | 8 hours | 2 | 2 | Non-dispersive infrared (NDIR) spectroscopy |
| CO, mg/m³ | 1 hour | 4 | 4 | As above |
| Ammonia, µg/m³ | Annual | 100 | 100 | Chemiluminescence; indophenol blue |
| Ammonia, µg/m³ | 24 hours | 400 | 400 | As above |
| Benzene, µg/m³ | Annual | 5 | 5 | GC-based continuous analyser; adsorption and desorption followed by GC |
| Benzo(a)pyrene, particulate phase only, ng/m³ | Annual | 1 | 1 | Solvent extraction followed by HPLC or GC |
| Arsenic, ng/m³ | Annual | 6 | 6 | AAS or ICP after sampling on EPM 2000 or equivalent filter |
| Nickel, ng/m³ | Annual | 20 | 20 | AAS or ICP after sampling on EPM 2000 or equivalent filter |
Watch the units: CO is in mg/m³, benzo(a)pyrene, arsenic and nickel are in ng/m³, and the rest are in µg/m³.
The notification's footnotes matter as much as the numbers:
- An annual standard is the annual arithmetic mean of at least 104 measurements a year at a site, taken twice a week, 24-hourly, at uniform intervals.
- The 24-hour, 8-hour and 1-hour standards shall be met 98 % of the time in a year. They may be exceeded 2 % of the time, but not on two consecutive days of monitoring.
- When results exceed the limits on two consecutive monitoring days, that is adequate reason to start regular or continuous monitoring and further investigation.
Which ambient air monitoring parameters should a lab measure?
Under CPCB's National Air Quality Monitoring Programme (NAMP), four pollutants are monitored at all locations: SO2, NO2, PM10 and PM2.5. CPCB also records wind speed and direction, relative humidity and temperature, because results mean little without the weather that produced them.
Projects and industries often need more of the 12 NAAQS parameters, such as ozone, CO, ammonia, lead or benzene. The list is set by the consent, the environmental clearance conditions or the customer's monitoring plan. Agree it in writing before the field team leaves.
What is the ambient air quality monitoring frequency under CPCB guidelines?
Under NAMP, CPCB monitors for 24 hours, twice a week, to get 104 observations in a year. Gaseous pollutants are sampled 4-hourly and particulate matter 8-hourly, and PM2.5 is sampled on a 24-hour basis.
This ambient air quality monitoring frequency is what an annual NAAQS comparison needs. A one-off or short campaign can only be compared with the 24-hour, 8-hour or 1-hour standards. Never compare a single 24-hour PM10 result with the annual limit of 60 µg/m³, and never call one result an annual average.
How is PM10 measured by the gravimetric method?
CPCB's Guidelines for the Measurement of Ambient Air Pollutants, Volume I, describe the gravimetric method. Air is drawn through a size-selective inlet and a 20.3 × 25.4 cm glass fibre filter, typically at 1132 L/min. PM10 is the weight gained by the filter divided by the volume of air sampled.
The main steps from the CPCB protocol:
- Inspect the filter for pinholes and mark its identification number.
- Condition the filter for 24 hours in a room at 20 to 30 °C and 40 to 50 % relative humidity, or in an airtight desiccator, then weigh it (initial weight).
- Load the filter, record the flow rate and the elapsed-time meter, and sample, commonly for 8 or 24 hours, reading the flow rate hourly.
- Remove the filter into a clean envelope or container, and record total time and average flow.
- Condition the exposed filter again for 24 hours and weigh it (final weight).
- Calculate the concentration: C (µg/m³) = (Wf − Wi) × 10⁶ / V, with weights in grams and V in m³.
The CPCB protocol cites IS 5182 (Part 23) and calls for periodic calibration of the sampler with an orifice transfer standard. Field and lab blanks belong in your quality control.
Worked example: PM10 from filter weights
Here is a 24-hour run, worked through step by step.
| Input | Value |
|---|---|
| Filter weight before sampling (Wi) | 2512.4 mg |
| Filter weight after sampling (Wf) | 2650.8 mg |
| Average flow rate | 1.12 m³/min |
| Sampling time | 1440 min |
- Volume of air: V = 1.12 m³/min × 1440 min = 1612.8 m³.
- Mass collected: 2650.8 − 2512.4 = 138.4 mg = 138,400 µg.
- Concentration: 138,400 µg / 1612.8 m³ = 85.8 µg/m³, reported as 86 µg/m³.
- Comparison: 86 µg/m³ is within the 24-hour PM10 limit of 100 µg/m³ for both area categories.
Two unit traps catch new analysts. If weights are entered in grams in a formula that expects milligrams, the result is a thousand times too small and will always "pass". If the flow is in L/min and treated as m³/min, the result is a thousand times too small as well. Write the expected unit beside every input on the worksheet.
Where should an ambient air sampler be sited?
Site the sampler where it represents the area, away from the direct influence of a single source. CPCB's PM2.5 protocol in Volume I gives these siting rules:
- Inlet height 3 to 10 m above ground, at a suitable distance from any direct pollution source, including traffic.
- Distance to an obstacle such as a building more than twice the obstacle's height above the sampler.
- Unrestricted airflow in three of the four quadrants.
- More than 20 m from trees.
- At least 2 m between collocated samplers.
Record the location with coordinates, photographs and a sketch, so that a later campaign can return to the same point.
How should ambient air monitoring results be reported?
A good ambient air report lets the reader check every number against the standard. Include:
- Customer, project and the purpose of monitoring.
- Each location: name, coordinates and area category.
- Start and end date and time, duration, and the averaging time of each result.
- Sampler ID, its calibration status, average flow rate and air volume.
- Filter or absorber IDs and the method for each parameter.
- Weather during sampling.
- Results in NAAQS units, beside the limit for the same averaging time and area category.
- A clear note where a result exceeds a limit, and whether the next monitoring day also exceeded.
For several locations in one campaign, a combined report with parameters against locations is easier to read than one report per filter. See our guide to NABL test report format for the general report requirements.
MoEF&CC recognises environmental laboratories under Sections 12 and 13 of the Environment (Protection) Act, 1986, and customers may also want NABL accreditation for the parameters. Check what the consent or clearance requires before you quote.
Where Tanush LIMS fits
Tanush LIMS includes NAAQS 2009 in its library of Indian standards. Like all its standards, it loads as a draft that your technical manager checks and activates. At inward, staff choose the "Area category" and the "Averaging time" (pre-filled as 24 hours), so the right limit is applied.
In Formula mode, the PM10 worksheet takes filter weights in mg, flow in m³/min and time in minutes. The example above gives 86 µg/m³, shown against "≤ 100" as "Complies". Colorimetric methods such as SO2, NO2, NH3 and ozone take a calibration factor once per worksheet. Several samples from one campaign can go out as one combined report. See Tanush LIMS for air, stack and noise labs or book a demo.
Frequently asked questions
What is NAAQS 2009?
NAAQS 2009 is the set of National Ambient Air Quality Standards notified by CPCB on 18 November 2009 under the Air Act, 1981. It sets limits for 12 pollutants: SO2, NO2, PM10, PM2.5, ozone, lead, CO, ammonia, benzene, benzo(a)pyrene, arsenic and nickel.
What are the CPCB ambient air quality monitoring guidelines for frequency?
Under the National Air Quality Monitoring Programme, CPCB samples for 24 hours, twice a week, to get 104 observations a year. The NAAQS annual standards assume at least 104 such measurements at a site.
Can I compare one 24-hour result with the annual standard?
No. A single 24-hour result can only be compared with the 24-hour standard. The annual standard is the mean of at least 104 measurements taken through the year.
What is the PM10 formula?
CPCB gives C (µg/m³) = (Wf − Wi) × 10⁶ / V, where Wf and Wi are the final and initial filter weights in grams and V is the volume of air sampled in m³. Volume is the average flow rate multiplied by the sampling time.
Which ambient air monitoring parameters are compulsory?
CPCB monitors SO2, NO2, PM10 and PM2.5 at all NAMP locations. For an industry or project, the parameters are those in its consent, environmental clearance or monitoring plan, chosen from the 12 NAAQS pollutants.
Has NAAQS 2009 been revised?
As of October 2026 we have not found a notification replacing it, and CPCB's website still publishes it. Check CPCB's website before relying on it for a new contract.
Sources
- CPCB, National Ambient Air Quality Standards, Notification No. B-29016/20/90/PCI-L, 18 November 2009 (Gazette of India, Extraordinary, Part III, Section 4).
- CPCB, About the National Air Quality Monitoring Programme.
- CPCB, Technical guidelines and SOPs, including Guidelines for the Measurement of Ambient Air Pollutants, Volume I.
- MoEF&CC, Report in OA No. 627 of 2019 before the National Green Tribunal (recognition of environmental laboratories under Sections 12 and 13 of the E(P) Act).